What to do when things go wrong with the QC or you get an error message

There is a set procedure that must be followed if any QC results fall outside of the stated QC ranges. The QC ranges can be found on the relevant QC worksheets located in the Operators Manual; or displayed electronically with their results on the analyser screen; or printed out automatically after analysis. As with all QC print outs, each of the results will be printed alongside their relevant QC ranges. Any QC results that fail will be marked with an asterisk or question mark.

 

NOTE: the same procedure described below applies to other Error Messages as well, eg, analyser error messages or sample error messages. Please document EVERYTHING, even if only to say that you asked the POCT Coordinator to sort out the problem.

 

PLEASE READ THESE IMPORTANT POINTS:

 

Do not ignore error messages. They are there for a good reason.

 

If a QC sample fails, do not use the analyser for patient testing until you have a valid QC result. You cannot assume the patient results will be correct if the QC is wrong. 

 

If a repeat QC analysis gives the same error, do not keep running QC samples until you get one that works, thinking that the problem has resolved itself. It hasn't, trust me on this.... 

 

If you can't fix the problem, please seek help via the POCT Coordinator.

 


Remember: document all of your actions.


Resolving the error. 

 

Please work your way systematically through the list below. Your resource is the Point of Care Coordinator. Please contact him for any assistance you may need.

 

1.

Error messages- please consult the Operators Manual for a list of common error messages and how to fix them.

 

2.

Repeat the QC analysis- (this will make sure that the result is not a random error).

 

3.

Check that your repeat QC sample is totally thawed if frozen. Make sure it is well-mixed- this is very important. All samples must be well mixed before analysis.

 

4.

Check your technique- eg, if it is a pO2 or a pO2 blood gas QC problem, are you making sure your sample remains as unexposed to air as possible before analysis? If a blood gas sample, is it clotted? Have you ejected the first few drops before analysis? Is the analyser correctly calibrated?

 

5.

Check for an:

 

incorrectly selected QC level- check both the actual sample and the analyser QC file into which you are placing the results. If you are writing onto a QC worksheet check if it is the appropriate one for your sample. 

 

incorrect lot number- the lot number of the sample must always match that inscribed on the QC worksheet. Contact the POCT Coordinator if there is a mis-match.

 

out-of-date QC sample- contact the POCT Coordinator if the QC is out-of-date.

 

6.

Review the analyser operation for procedural errors- contact the POCT Coordinator for a refresher course or revision.

 

7.

Check for faulty or expired reagents- contact the POCT Coordinator if reagents are out-of-date.

 

8.

Check for calibration problems and recalibrate if necessary- check the QC Error Log Sheet in the Operator's Manual for past documented errors.

 

9.

Check for any analyser faults- contact the POCT Coordinator if you suspect there is an analyser fault.